Suffix:
<ri>
.
1..n
Selects the measurement range.
Parameters:
<SweepTime> Sweep time.
The range depends on the ratios of the span to the RBW and
the RBW to the VBW. Refer to the data sheet for more informa-
tion.
Example:
LIST:RANG2:SWE:TIME 2MS
Manual operation: See "Sweep Time" on page 205
[SENSe:]LIST:RANGe<ri>:SWEep:TIME:AUTO <State>
This command turns automatic selection of the sweep time for a spurious emission
measurement range on and off.
Suffix:
<ri>
.
1..n
Selects the measurement range.
Parameters:
<State> ON | OFF | 0 | 1
*RST: 1
Example:
LIST:RANG2:SWE:TIME:AUTO ON
Manual operation: See "Sweep Time Mode" on page 205
[SENSe:]LIST:RANGe<ri>:TRANsducer <Transducer>
This command selects a transducer factor for a spurious emission measurement
range.
Note that
●
the transducer must cover at least the span of the range
●
the x-axis has to be linear
●
the unit has to be dB
Suffix:
<ri>
.
1..n
Selects the measurement range.
Parameters:
<Transducer> String containing the transducer file name, including the path
information.
Example:
LIST:RANG2:TRAN 'C:\TRANS\MYTRANS'
Manual operation: See "Transducer" on page 207