*RST: 0
Example:
CALC:ESP:PSE:DET ON
CALC:PSE:SUBR 10
Includes up to 10 peaks per range in the list.
Manual operation: See "Details" on page 209
CALCulate<n>:PSEarch:MARGin <Threshold>
CALCulate<n>:PEAKsearch:MARGin <Margin>
This command defines the threshold of the list evaluation.
Suffix:
<n>
.
Window
Parameters:
<Margin> Range: -200 to 200
Default unit: dB
Example:
CALC:PSE:MARG 100
Sets the threshold to 100 dB.
Manual operation: See "Margin" on page 209
CALCulate<n>:PSEarch:PSHow <State>
CALCulate<n>:PEAKsearch:PSHow <State>
This command turns the peak labels in the diagram on and off.
Peak labels are blue squares.
Suffix:
<n>
.
Window
Parameters:
<State> ON | OFF | 1 | 0
*RST: 0
Example:
CALC:PSE:PSH ON
Marks all peaks with blue squares.
Manual operation: See "Show Peaks" on page 208
CALCulate<n>:PSEarch:SUBRanges <NumberPeaks>
CALCulate<n>:PEAKsearch:SUBRanges <NumberPeaks>
This command defines the number of peaks included in the peak list.
After this number of peaks has been found, the R&S FPL1000 stops the peak search
and continues the search in the next measurement range.
Suffix:
<n>
.
Window