User Manual
Remote Commands
R&S
®
FPL1000
699User Manual 1178.3370.02 ─ 09
Parameters:
<NumberPeaks> Range: 1 to 50
*RST: 25
Example:
CALC:PSE:SUBR 10
Sets 10 peaks per range to be stored in the list.
Manual operation: See "Peaks per Range" on page 209
11.6.8.4 Adjusting the X-Axis to the Range Definitions
[SENSe:]LIST:XADJust
Sets the x-axis range for the spurious emission measurement from the start frequency
of the first sweep range to the stop frequency of the last sweep range.
Example:
SENS:LIST:XADJ
Usage: Event
11.6.8.5
Performing a Spurious Measurement
The following commands are required to perform a Spurious measurement:
SENS:SWE:MODE LIST, see [SENSe:]SWEep:MODE on page 661
INITiate<n>[:IMMediate] on page 629, see Chapter 11.6.1, "Performing Mea-
surements", on page 626
11.6.8.6 Retrieving and Saving Settings and Results
The following commands analyze and retrieve measurement results for Spurious mea-
surements.
Useful commands for spurious emission measurements described elsewhere
●
CALCulate<n>:LIMit<li>:FAIL? on page 914
●
TRACe<n>[:DATA] on page 832
●
TRACe<n>[:DATA]:MEMory? on page 833
●
TRACe<n>[:DATA]:X? on page 834
11.6.8.7 Programming Example: Spurious Emissions Measurement
In the following example, the Spurious Emissions measurement is configured by defin-
ing ranges and parameters to create the following sweep list.
Note that this example is primarily meant to demonstrate the remote control com-
mands, it does not necessarily reflect a useful measurement task.
//------------Preparing the measurement-------------
*RST
//Resets the instrument
Configuring and Performing Measurements