The gate length applies to all traces.
Suffix:
<t>
.
irrelevant
Parameters:
<Length> Range: 100 ns to 1000 s
*RST: 2 ms
Default unit: s
Example:
SWE:EGAT:TRAC:PER 5ms
Defines the period for gated triggering to 5 ms.
Manual operation: See "Period" on page 219
[SENSe:]SWEep:EGATe:TRACe<t>:STARt<gr> <Time>
This command defines the start time for a gate range.
Suffix:
<t>
.
Trace
<gr> 1..n
gate range
Parameters:
<Time> The value range depends on the gate period you have set for
the selected trace with [SENSe:]SWEep:EGATe:TRACe<t>:
PERiod. The following rules apply:
• the start time may not be higher than the length of the gate
• the start time may not be lower than the stop time of the gate
range of a lower order
The reset values depend on the gate range.
• for gate range 1, the start time is 0 ms
• for gate range 3, the start time is 2 ms
• for gate range 5, the start time is 4 ms
Default unit: s
Example:
SWE:EGAT:TRAC1:STAR1 3ms
Sets the Starting point for range 1 on trace 1 at 3 ms.
Manual operation: See "Range <x> Start/Stop" on page 219
[SENSe:]SWEep:EGATe:TRACe<t>[:STATe<gr>] <State>
This command includes or excludes a gate range for a particular trace.
Suffix:
<t>
.
Trace
<gr> gate range
Parameters:
<State> ON | OFF | 0 | 1
Configuring and Performing Measurements