CALCulate<n>:STATistics:SCALe:X:RLEVel <RefLevel>
This command sets the reference level for statistical measurements. The effects are
identical to DISPlay[:WINDow<n>]:TRACe<t>:Y[:SCALe]:RLEVel.
Note that in case of statistical measurements the reference level applies to the x-axis.
Suffix:
<n>
.
Window
Parameters:
<RefLevel> The unit is variable.
If a reference level offset is included, the range is adjusted by
that offset.
Range: -130 dBm to 30 dBm
*RST: 0 dBm
Default unit: dBm
Example:
CALC:STAT:SCAL:X:RLEV -60dBm
Manual operation: See "Ref Level" on page 221
CALCulate<n>:STATistics:SCALe:Y:LOWer <Magnitude>
This command defines the lower vertical limit of the diagram.
Suffix:
<n>
.
Window
Parameters:
<Magnitude> The number is a statistical value and therefore dimensionless.
Range: 1E-9 to 0.1
*RST: 1E-6
Example:
CALC:STAT:SCAL:Y:LOW 0.001
Manual operation: See "Y-Max/ Y-Min" on page 221
CALCulate<n>:STATistics:SCALe:Y:UNIT <Unit>
This command selects the unit of the y-axis.
Suffix:
<n>
.
Window
Parameters:
<Unit> PCT | ABS
*RST: ABS
Example:
CALC:STAT:SCAL:Y:UNIT PCT
Sets the percentage scale.
Manual operation: See "Y-Unit" on page 221