CALC:STAT:SCAL:Y:UNIT PCT
//Displays percentage values on y-axis scale
11.6.10 Measuring the Time Domain Power
All remote control commands specific to time domain power measurements are descri-
bed here.
Configuring the Measurement...............................................................................711
Performing a Time Domain Power Measurement................................................. 714
Retrieving Measurement Results..........................................................................714
Programming Example: Time Domain Power....................................................... 719
11.6.10.1 Configuring the Measurement
The following remote commands measure the time domain power.
Useful commands for time domain power measurements described elsewhere
CALCulate<n>:MARKer<m>:X:SLIMits:LEFT
CALCulate<n>:MARKer<m>:X:SLIMits:RIGHt
CALCulate<n>:MARKer<m>:X:SLIMits[:STATe]
Remote commands exclusive to time domain power measurements
CALCulate<n>:MARKer<m>:FUNCtion:SUMMary:AOFF....................................................711
CALCulate<n>:MARKer<m>:FUNCtion:SUMMary:AVERage.............................................. 711
CALCulate<n>:MARKer<m>:FUNCtion:SUMMary:PHOLd................................................. 712
CALCulate<n>:MARKer<m>:FUNCtion:SUMMary[:STATe].................................................712
CALCulate<n>:MARKer<m>:FUNCtion:SUMMary:MEAN[:STATe].......................................713
CALCulate<n>:MARKer<m>:FUNCtion:SUMMary:PPEak[:STATe]...................................... 713
CALCulate<n>:MARKer<m>:FUNCtion:SUMMary:RMS[:STATe].........................................713
CALCulate<n>:MARKer<m>:FUNCtion:SUMMary:SDEViation[:STATe]................................714
CALCulate<n>:MARKer<m>:FUNCtion:SUMMary:AOFF
This command turns all time domain power evaluation modes off.
Suffix:
<n>
.
Window
<m> Marker
CALCulate<n>:MARKer<m>:FUNCtion:SUMMary:AVERage <State>
This command switches on or off averaging for the active power measurement in zero
span in the window specified by the suffix <n>. If activated, a time domain value is cal-
culated from the trace after each sweep; in the end, all values are averaged to calcu-
late the final result.
Configuring and Performing Measurements