User Manual
Remote Commands
R&S
®
FPL1000
732User Manual 1178.3370.02 ─ 09
Parameters:
<State>
CALCulate<n>:DELTamarker<m>:FUNCtion:FMEasurement:DWELl <Time>
CALCulate<n>:MARKer<m>:FUNCtion:FMEasurement:DWELl <Time>
This command defines the dwell time during the final measurement.
Suffix:
<n>
.
irrelevant
<m> irrelevant
Parameters:
<Time> Range: 100 us to 100 s
*RST: 1 s
Default unit: s
Manual operation: See "Dwell Time" on page 267
11.6.14.4 Configuring EMI Limit Lines
The commands required to define limit lines for EMI measurements are described in
Chapter 11.8.9, "Configuring Display Lines", on page 898.
11.6.14.5 Controlling LISN
The commands required to control a LISN are described here.
This feature requires the optional additional interfaces (R&S FPL1-B5).
INPut<ip>:LISN:FILTer:HPASs[:STATe].............................................................................732
INPut<ip>:LISN:PHASe.................................................................................................. 733
INPut<ip>:LISN[:TYPE]...................................................................................................733
INPut<ip>:LISN:FILTer:HPASs[:STATe] <State>
This command turns the 150 kHz highpass filter for the ENV216 network on and off.
Suffix:
<ip>
.
1 | 2
irrelevant
Parameters:
<State> ON | OFF | 0 | 1
OFF | 0
Switches the function off
ON | 1
Switches the function on
*RST: 0
Configuring and Performing Measurements