<RFAttenuation> Defines the RF attenuation for a list entry.
Range: 0 to 70
Increment: 1
Default unit: dB
<FilterType> Selects the filter type for a list entry. For more information see
[SENSe:]BANDwidth[:RESolution]:TYPE.
<RBW> Defines the resolution bandwidth for a list entry.
<VBW> Defines the measurement time for a list entry.
<MeasTime> Defines the measurement time for a list entry.
Range: 1 µs to 16000 s
Default unit: s
<TriggerLevel> The trigger level must be 0.
<PowerLevel> Power level for each frequency included in the measurement.
The command returns up to 3 power levels for each frequency,
depending on the number of evaluation modes you have turned
on with [SENSe:]LIST:POWer:SET.
The result is a list of floating point values separated by commas.
The unit depends on CALCulate<n>:UNIT:POWer.
[SENSe:]LIST:POWer:SET <State>, <State>, <State>, <TriggerSource>,
<TriggerSlope>, <TriggerOffset>, <GateLength>
This command defines global List Evaluation parameters.
These parameters are valid for every frequency you want to measure.
The state of the first three parameters (<PeakPower>, <RMSPower> and <AVG-
Power>) define the number of results for each frequency in the list.
Note that you have to set the trigger level after sending this command.
Parameters:
<State> ON | OFF | 0 | 1
Turns peak power evaluation on and off.
*RST: 1
<State> ON | OFF | 0 | 1
Turns RMS power evaluation on and off.
*RST: 0
<State> ON | OFF | 0 | 1
Turns average power evaluation on and off.
*RST: 0
<TriggerSource> IMMediate | LINE | EXTernal | VIDeo | RFPower | IFPower |
RFPower | EXT2 | EXT3 | EXT4 | LXI
Selects a trigger source.
Configuring and Performing Measurements