User Manual

Remote Commands
R&S
®
FPL1000
764User Manual 1178.3370.02 ─ 09
[SENSe:]ADJust:ALL......................................................................................................764
[SENSe:]ADJust:CONFigure:DURation............................................................................ 764
[SENSe:]ADJust:CONFigure:DURation:MODE..................................................................765
[SENSe:]ADJust:CONFigure:HYSTeresis:LOWer.............................................................. 765
[SENSe:]ADJust:CONFigure:HYSTeresis:UPPer...............................................................765
[SENSe:]ADJust:CONFigure:TRIGger..............................................................................766
[SENSe:]ADJust:FREQuency..........................................................................................766
[SENSe:]ADJust:LEVel................................................................................................... 766
[SENSe:]ADJust:ALL
This command initiates a measurement to determine and set the ideal settings for the
current task automatically (only once for the current measurement).
This includes:
Center frequency
Reference level
Example:
ADJ:ALL
Manual operation: See "Adjusting all Determinable Settings Automatically (Auto
All)" on page 344
[SENSe:]ADJust:CONFigure:DURation <Duration>
In order to determine the ideal reference level, the R&S FPL1000 performs a measure-
ment on the current input data. This command defines the length of the measurement
if [SENSe:]ADJust:CONFigure:DURation:MODE is set to MANual.
Parameters:
<Duration> Numeric value in seconds
Range: 0.001 to 16000.0
*RST: 0.001
Default unit: s
Example:
ADJ:CONF:DUR:MODE MAN
Selects manual definition of the measurement length.
ADJ:CONF:LEV:DUR 5ms
Length of the measurement is 5 ms.
Manual operation: See "Changing the Automatic Measurement Time (Meastime
Manual)" on page 345
Setting Basic Spectrum RF Measurement Parameters