User Manual
Remote Commands
R&S
®
FPL1000
768User Manual 1178.3370.02 ─ 09
Parameters:
<Bandwidth> Range: 10 Hz to Max span
*RST: (= span/10 on activating the function)
Default unit: Hz
Manual operation: See "Signal Tracking" on page 306
CALCulate<n>:MARKer<m>:FUNCtion:STRack:THReshold <Level>
This command defines the threshold level for the signal tracking process.
Note that you have to turn on signal tracking before you can use the command.
Suffix:
<n>
.
irrelevant
<m> irrelevant
Parameters:
<Level> The unit depends on CALCulate<n>:UNIT:POWer.
Range: -130 dBm to 30 dBm
*RST: -120 dBm
Default unit: DBM
Manual operation: See "Signal Tracking" on page 306
CALCulate<n>:MARKer<m>:FUNCtion:STRack:TRACe <TraceNumber>
This command selects the trace on which the largest signal is searched for.
Suffix:
<n>
.
irrelevant
<m> irrelevant
Parameters:
<TraceNumber> 1 to 6
Range: 1 to 6
*RST: 1
Manual operation: See "Signal Tracking" on page 306
11.8.2 Configuring Bandwidth and Sweep Settings
The commands required to configure the bandwidth, sweep and filter settings in a
remote environment are described here. The tasks for manual operation are described
in Chapter 7.6, "Bandwidth, Filter and Sweep Configuration", on page 316.
● Configuring the Bandwidth and Filter....................................................................769
● Configuring the Sweep..........................................................................................772
Setting Basic Spectrum RF Measurement Parameters