User Manual
Remote Commands
R&S
®
FPL1000
771User Manual 1178.3370.02 ─ 09
Manual operation: See "Filter Type" on page 266
See "Res BW CISPR" on page 268
See "Res BW MIL" on page 268
[SENSe:]BWIDth:VIDeo <Bandwidth>
[SENSe:]BANDwidth:VIDeo <Bandwidth>
This command defines the video bandwidth.
The command decouples the video bandwidth from the resolution bandwidths.
Parameters:
<Bandwidth> refer to data sheet
*RST: AUTO is set to ON
Default unit: HZ
Example:
BAND:VID 10 kHz
Manual operation: See "VBW" on page 324
[SENSe:]BWIDth:VIDeo:AUTO <State>
[SENSe:]BANDwidth:VIDeo:AUTO <State>
This command couples and decouples the video bandwidth to the resolution band-
width.
Parameters:
<State> ON | OFF | 0 | 1
*RST: 1
Example:
BAND:VID:AUTO OFF
Manual operation: See "VBW" on page 324
See "RBW/VBW" on page 325
See "Default Coupling" on page 326
[SENSe:]BWIDth:VIDeo:RATio <Ratio>
[SENSe:]BANDwidth:VIDeo:RATio <Ratio>
This command defines the coupling ratio of the video bandwidth to the resolution band-
width (RBW/VBW).
Parameters:
<Ratio> Range: 0,001 to 1000
*RST: 1
Example:
BAND:VID:RAT 3
Sets the coupling of video bandwidth to video bandwidth =
3*resolution bandwidth
Manual operation: See "RBW/VBW" on page 325
Setting Basic Spectrum RF Measurement Parameters