Parameters:
<SweepCount> When you set a sweep count of 0 or 1, the R&S FPL1000 per-
forms one single sweep in single sweep mode.
In continuous sweep mode, if the sweep count is set to 0, a
moving average over 10 sweeps is performed.
Range: 0 to 200000
*RST: 0
Example:
SWE:COUN 64
Sets the number of sweeps to 64.
INIT:CONT OFF
Switches to single sweep mode.
INIT;*WAI
Starts a sweep and waits for its end.
Manual operation: See "Sweep/Average Count" on page 326
[SENSe:]SWEep:DURation? <Time>
This command provides an estimation of the total time required to capture the data and
process it. This time span may be considerably longer than the actual sweep time (see
[SENSe:]SWEep:TIME on page 775).
Tip: To determine the necessary timeout for data capturing in a remote control pro-
gram, double the estimated time and add 1 second.
Return values:
<Time>
Example:
SWE:TIME 1s
SWE:DUR?
Reply:
27.9734842578
Usage: Query only
Manual operation: See "Sweep Time" on page 324
See "Data capturing takes too long" on page 1028
[SENSe:]SWEep:OPTimize <Mode>
In FFT mode, several FFT analysis steps are required to cover the entire measurement
span. The span which is covered by one FFT analysis step is called subspan. The sub-
span cannot be defined directly, but it can be optimized according to measurement
requirements.
Setting Basic Spectrum RF Measurement Parameters