User Manual

Remote Commands
R&S
®
FPL1000
775User Manual 1178.3370.02 ─ 09
Manual operation: See "Sweep Points" on page 326
[SENSe:]SWEep:TIME <Time>
This command defines the sweep time. It automatically decouples the time from any
other settings.
In the Spectrum application, the command decouples the sweep time from the span
and resolution and video bandwidths. Note that this command queries only the time
required to capture the data, not to process it. To obtain an estimation of the total cap-
ture and processing time, use the [SENSe:]SWEep:DURation? command.
Parameters:
<Time> refer to data sheet
*RST: depends on current settings (determined automati-
cally)
Default unit: S
Manual operation: See "Sweep Time" on page 138
See "Sweep Time" on page 324
See "Sweep Time" on page 336
[SENSe:]SWEep:TIME:AUTO <State>
This command couples and decouples the sweep time to the span and the resolution
and video bandwidths.
Parameters:
<State> ON | OFF | 0 | 1
*RST: 1
Example:
SWE:TIME:AUTO ON
Activates automatic sweep time.
Manual operation: See "Harmonic Sweep Time" on page 236
See "Sweep Time" on page 324
See "Default Coupling" on page 326
[SENSe:]SWEep:TYPE:USED
This command queries the sweep type if you have turned on automatic selection of the
sweep type.
Return values:
<Type> SWE
Normal sweep
FFT
FFT mode
Setting Basic Spectrum RF Measurement Parameters