780User Manual 1178.3370.02 ─ 09
Example:
DISP:TRAC:Y 110dB
Manual operation: See "Range" on page 314
DISPlay[:WINDow<n>]:TRACe<t>:Y[:SCALe]:AUTO ONCE
Automatic scaling of the y-axis is performed once, then switched off again (for all
traces).
Suffix:
<n>
.
Window
<t> irrelevant
Manual operation: See "Auto Scale Once" on page 315
DISPlay[:WINDow<n>][:SUBWindow<w>]:TRACe<t>:Y[:SCALe]:MODE
<Mode>
This command selects the type of scaling of the y-axis (for all traces).
When the display update during remote control is off, this command has no immediate
effect.
Suffix:
<n>
.
Window
<w> subwindow
<t> irrelevant
Parameters:
<Mode> ABSolute
absolute scaling of the y-axis
RELative
relative scaling of the y-axis
*RST: ABSolute
Example:
DISP:TRAC:Y:MODE REL
Manual operation: See "Scaling" on page 315
DISPlay[:WINDow<n>]:TRACe<t>:Y[:SCALe]:RPOSition <Position>
This command defines the vertical position of the reference level on the display grid
(for all traces).
The R&S FPL1000 adjusts the scaling of the y-axis accordingly.
For measurements with the optional tracking generator, the command defines the posi-
tion of the reference line.
Suffix:
<n>
.
Window
<t> irrelevant
Setting Basic Spectrum RF Measurement Parameters