User Manual
Remote Commands
R&S
®
FPL1000
781User Manual 1178.3370.02 ─ 09
Parameters:
<Position> 0 PCT corresponds to the lower display border, 100 percent cor-
responds to the upper display border.
*RST: frequency display: 90 PCT; time display: 50 PCT;
AF spectrum display (K7): 100 PCT;
Default unit: PCT
Example:
DISP:TRAC:Y:RPOS 50PCT
Manual operation: See "Reference Position" on page 294
See "Ref Level Position" on page 314
DISPlay[:WINDow<n>][:SUBWindow<w>]:TRACe<t>:Y:SPACing <ScalingType>
This command selects the scaling of the y-axis (for all traces, <t> is irrelevant).
Suffix:
<n>
.
Window
<w> subwindow
<t> Trace
Parameters:
<ScalingType> LOGarithmic
Logarithmic scaling.
LINear
Linear scaling in %.
LDB
Linear scaling in the specified unit.
PERCent
Linear scaling in %.
*RST: LOGarithmic
Example:
DISP:TRAC:Y:SPAC LIN
Selects linear scaling in %.
Manual operation: See "Scaling" on page 315
11.8.4 Configuring Triggered and Gated Measurements
The commands required to configure a triggered or gated measurement in a remote
environment are described here.
The tasks for manual operation are described in Chapter 7.7, "Trigger and Gate Con-
figuration", on page 331.
Setting Basic Spectrum RF Measurement Parameters