User Manual
Remote Commands
R&S
®
FPL1000
782User Manual 1178.3370.02 ─ 09
*OPC should be used after requesting data. This will hold off any subsequent changes
to the selected trigger source, until after the sweep is completed and the data is
returned.
● Configuring the Triggering Conditions...................................................................782
● Configuring Gated Measurements........................................................................ 786
11.8.4.1 Configuring the Triggering Conditions
The following commands are required to configure a triggered measurement.
TRIGger[:SEQuence]:DTIMe...........................................................................................782
TRIGger[:SEQuence]:HOLDoff[:TIME]..............................................................................782
TRIGger[:SEQuence]:IFPower:HOLDoff...........................................................................783
TRIGger[:SEQuence]:IFPower:HYSTeresis.......................................................................783
TRIGger[:SEQuence]:LEVel[:EXTernal<port>]...................................................................784
TRIGger[:SEQuence]:LEVel:IFPower............................................................................... 784
TRIGger[:SEQuence]:LEVel:IQPower...............................................................................784
TRIGger[:SEQuence]:SLOPe.......................................................................................... 785
TRIGger[:SEQuence]:SOURce........................................................................................785
TRIGger[:SEQuence]:TIME:RINTerval..............................................................................786
TRIGger[:SEQuence]:DTIMe <DropoutTime>
Defines the time the input signal must stay below the trigger level before a trigger is
detected again.
Parameters:
<DropoutTime> Dropout time of the trigger.
Range: 0 s to 10.0 s
*RST: 0 s
Default unit: S
Manual operation: See "Drop-Out Time" on page 338
TRIGger[:SEQuence]:HOLDoff[:TIME] <Offset>
Defines the time offset between the trigger event and the start of the sweep.
A negative offset is possible for time domain measurements.
For the trigger sources "External" or "IF Power", a common input signal is used for both
trigger and gate. Therefore, changes to the gate delay will affect the trigger offset as
well.
Setting Basic Spectrum RF Measurement Parameters