User Manual
Remote Commands
R&S
®
FPL1000
783User Manual 1178.3370.02 ─ 09
Parameters:
<Offset> For measurements in the frequency domain, the range is 0 s to
30 s.
For measurements in the time domain, the range is the negative
sweep time to 30 s.
*RST: 0 s
Default unit: S
Example:
TRIG:HOLD 500us
Manual operation: See "Trigger Offset" on page 338
TRIGger[:SEQuence]:IFPower:HOLDoff <Period>
This command defines the holding time before the next trigger event.
Note that this command can be used for any trigger source, not just IF Power
(despite the legacy keyword).
Note: If you perform gated measurements in combination with the IF Power trigger, the
R&S FPL1000 ignores the holding time for frequency sweep, FFT sweep, zero span
and I/Q data measurements.
Parameters:
<Period> Range: 0 s to 10 s
*RST: 0 s
Default unit: S
Example:
TRIG:SOUR EXT
Sets an external trigger source.
TRIG:IFP:HOLD 200 ns
Sets the holding time to 200 ns.
Manual operation: See "Trigger Holdoff" on page 339
TRIGger[:SEQuence]:IFPower:HYSTeresis <Hysteresis>
This command defines the trigger hysteresis, which is only available for "IF Power" trig-
ger sources.
Parameters:
<Hysteresis> Range: 3 dB to 50 dB
*RST: 3 dB
Default unit: DB
Example:
TRIG:SOUR IFP
Sets the IF power trigger source.
TRIG:IFP:HYST 10DB
Sets the hysteresis limit value.
Manual operation: See "Hysteresis" on page 338
Setting Basic Spectrum RF Measurement Parameters