TRIGger[:SEQuence]:LEVel[:EXTernal<port>] <TriggerLevel>
This command defines the level the external signal must exceed to cause a trigger
event.
Suffix:
<port>
.
Selects the trigger port.
1 = trigger port 1 (TRIG IN connector on rear panel)
2 = trigger port 2 (TRIG AUX connector on rear panel)
Parameters:
<TriggerLevel> For the R&S FPL1000, the external trigger level is always 1.4 V.
It cannot be changed.
*RST: 1.4 V
Manual operation: See "Trigger Level" on page 337
TRIGger[:SEQuence]:LEVel:IFPower
<TriggerLevel>
This command defines the power level at the third intermediate frequency that must be
exceeded to cause a trigger event.
Note that any RF attenuation or preamplification is considered when the trigger level is
analyzed. If defined, a reference level offset is also considered.
Parameters:
<TriggerLevel> For details on available trigger levels and trigger bandwidths see
the data sheet.
*RST: -10 dBm
Default unit: DBM
Example:
TRIG:LEV:IFP -30DBM
Manual operation: See "Trigger Level" on page 337
TRIGger[:SEQuence]:LEVel:IQPower <TriggerLevel>
This command defines the magnitude the I/Q data must exceed to cause a trigger
event.
Note that any RF attenuation or preamplification is considered when the trigger level is
analyzed. If defined, a reference level offset is also considered.
Parameters:
<TriggerLevel> Range: -130 dBm to 30 dBm
*RST: -20 dBm
Default unit: DBM
Example:
TRIG:LEV:IQP -30DBM
Manual operation: See "Trigger Level" on page 337