TRIGger[:SEQuence]:TIME:RINTerval <Interval>
This command defines the repetition interval for the time trigger.
Parameters:
<Interval> 2.0 ms to 5000
Range: 2 ms to 5000 s
*RST: 1.0 s
Default unit: S
Example:
TRIG:SOUR TIME
Selects the time trigger input for triggering.
TRIG:TIME:RINT 50
The sweep starts every 50 s.
Manual operation: See "Repetition Interval" on page 338
11.8.4.2 Configuring Gated Measurements
[SENSe:]SWEep:EGATe................................................................................................. 786
[SENSe:]SWEep:EGATe:HOLDoff....................................................................................787
[SENSe:]SWEep:EGATe:LENGth.....................................................................................787
[SENSe:]SWEep:EGATe:POLarity....................................................................................787
[SENSe:]SWEep:EGATe:SOURce................................................................................... 788
[SENSe:]SWEep:EGATe:TYPE........................................................................................788
[SENSe:]SWEep:EGATe <State>
This command turns gated measurements on and off.
The measurement ends when a particular number of measurement points has been
recorded.
(See [SENSe:]SWEep[:WINDow<n>]:POINts on page 774).
Performing gated measurements turns the squelch off.
Parameters:
<State> ON | OFF | 0 | 1
OFF | 0
Switches the function off
ON | 1
Switches the function on
Setting Basic Spectrum RF Measurement Parameters