User Manual

Remote Commands
R&S
®
FPL1000
801User Manual 1178.3370.02 ─ 09
Parameters:
<Frequency> Range: 5 kHz to 3 GHz
Increment: 0.1 Hz
*RST: 1 GHz
Default unit: HZ
Example:
SOUR:INT:FREQ 2 GHz
Manual operation: See "Frequency" on page 291
SOURce<si>:POWer[:LEVel][:IMMediate][:AMPLitude] <Amplitude>
Defines the output power of the internal generator.
Suffix:
<si>
.
irrelevant
Parameters:
<Amplitude> Range: -60 dBm to +10 dBm
Increment: 0.1 dB
*RST: -20 dBm
Default unit: DBM
Example:
SOUR:POW -30dBm
Manual operation: See "Level" on page 291
SOURce<si>:POWer[:LEVel][:IMMediate]:OFFSet <Offset>
This command defines a level offset for the external generator level. Thus, for exam-
ple, attenuators or amplifiers at the output of the external generator can be taken into
account for the setting.
Suffix:
<si>
.
irrelevant
Parameters:
<Offset> Range: -200 dB to +200 dB
*RST: 0dB
Default unit: DB
Example: //Define a level offset on the external generator
SOUR:POW:OFFS -10dB
Manual operation: See "Level Offset" on page 291
Tracking Generator Commands
The following commands are required to configure an internal tracking generator.
Useful commands for tracking generators described elsewhere:
OUTPut<up>[:STATe] on page 800
DISPlay[:WINDow<n>]:TRACe<t>:Y[:SCALe]:RPOSition on page 780
SOURce<si>:INTernal:FREQuency on page 800
Setting Basic Spectrum RF Measurement Parameters