FIXed
Fixed power level. Required for tracking generator.
SWEep
Increasing power level. Required for power sweep.
*RST: FIXed
Example:
SOUR:POW:MODE:SWE
SOUR:POW:STAR -50
SOUR:POW:STOP -10
Manual operation: See "Power Sweep" on page 98
See "Usage" on page 291
SOURce<si>:POWer:STARt
<StartLevel>
Defines the start power level of the internal generator signal for power sweeps. This
value must be lower than the stop level (see SOURce<si>:POWer:STOP
on page 806).
Suffix:
<si>
.
irrelevant
Parameters:
<StartLevel> Range: -60 to +10
*RST: -30
Default unit: DBM
Example:
SOUR:POW:MODE:SWE
SOUR:POW:STAR -50
SOUR:POW:STOP -10
Manual operation: See "Level Start / Level Stop" on page 297
SOURce<si>:POWer:STOP <StopLevel>
Defines the end power level of the internal generator signal for power sweeps. This
value must be higher than the start level (see SOURce<si>:POWer:STARt
on page 806).
Suffix:
<si>
.
irrelevant
Parameters:
<StopLevel> Range: -60 to +10
*RST: -20
Default unit: DBM
Example:
SOUR:POW:MODE:SWE
SOUR:POW:STAR -50
SOUR:POW:STOP -10
Manual operation: See "Level Start / Level Stop" on page 297