User Manual

Remote Commands
R&S
®
FPL1000
809User Manual 1178.3370.02 ─ 09
*RST: IF
Example:
OUTP:IF VID
Selects the video signal for the IF/VIDEO output connector.
Manual operation: See "Data Output" on page 300
OUTPut<up>:IF:IFFRequency <Frequency>
This command defines the frequency for the IF output of the R&S FPL1000. The IF fre-
quency of the signal is converted accordingly.
This command is available in the time domain and if the IF/VIDEO output is configured
for IF.
Suffix:
<up>
.
Parameters:
<Frequency> *RST: 50.0 MHz
Default unit: HZ
Manual operation: See "Data Output" on page 300
OUTPut<up>:UPORt:STATe <State>
This command toggles the control lines of the user ports for the AUX PORT connector.
This 9-pole SUB-D male connector is located on the rear panel of the R&S FPL1000.
Suffix:
<up>
.
irrelevant
Parameters:
<State> ON | OFF | 0 | 1
OFF | 0
User port is switched to INPut
ON | 1
User port is switched to OUTPut
Example:
OUTP:UPOR:STAT ON
OUTPut<up>:UPORt[:VALue] <Value>
This command sets the control lines of the user ports.
The assignment of the pin numbers to the bits is as follows:
Bit 7 6 5 4 3 2 1 0
Pin N/A N/A 5 3 4 7 6 2
Bits 7 and 6 are not assigned to pins and must always be 0.
The user port is written to with the given binary pattern.
Setting Basic Spectrum RF Measurement Parameters