User Manual
Remote Commands
R&S
®
FPL1000
819User Manual 1178.3370.02 ─ 09
Parameters:
<AverageCount> If you set an average count of 0 or 1, the application performs
one single sweep in single sweep mode.
In continuous sweep mode, if the average count is set to 0, a
moving average over 10 sweeps is performed.
Range: 0 to 200000
*RST: 0
Manual operation: See "Sweep/Average Count" on page 326
See "Average Count" on page 396
[SENSe:]AVERage<n>[:STATe<t>] <State>
This command turns averaging for a particular trace in a particular window on and off.
Suffix:
<n>
.
Window
<t> Trace
Parameters:
<State> ON | OFF | 1 | 0
[SENSe:]AVERage<n>:TYPE <Mode>
This command selects the trace averaging mode.
Suffix:
<n>
.
1..n
Window
Parameters:
<Mode> VIDeo
The logarithmic power values are averaged.
LINear
The power values are averaged before they are converted to
logarithmic values.
POWer
The power level values are converted into unit Watt prior to
averaging. After the averaging, the data is converted back into
its original unit.
*RST: VIDeo
Example:
AVER:TYPE LIN
Switches to linear average calculation.
Manual operation: See "Average Mode" on page 395
[SENSe:][WINDow<n>:]DETector<t>[:FUNCtion] <Detector>
Defines the trace detector to be used for trace analysis.
Setting Basic Spectrum RF Measurement Parameters