User Manual
Remote Commands
R&S
®
FPL1000
820User Manual 1178.3370.02 ─ 09
For details see "Mapping Samples to sweep Points with the Trace Detector"
on page 386.
For EMI measurements, the trace detector is used for the initial peak search only, not
for the final test. The detector for the final test is configured using CALCulate<n>:
MARKer<m>:FUNCtion:FMEasurement:DETector on page 730.
If the EMI (R&S FPL1-K54) measurement option is installed and the filter type "CISPR"
is selected, additional detectors are available, even if EMI measurement is not active.
For details see "Detectors and Dwell Time" on page 253.
Suffix:
<n>
.
Window
<t> Trace
Parameters:
<Detector> APEak
Autopeak
NEGative
Negative peak
POSitive
Positive peak
QPEak
Quasipeak (CISPR filter only)
SAMPle
First value detected per trace point
RMS
RMS value
AVERage
Average
CAVerage
CISPR Average (CISPR filter only)
CRMS
CISPR RMS (CISPR filter only)
*RST:
APEak
Example:
DET POS
Sets the detector to "positive peak".
Manual operation: See "Detector" on page 394
[SENSe:][WINDow<n>:]DETector<t>[:FUNCtion]:AUTO <State>
This command couples and decouples the detector to the trace mode.
Suffix:
<n>
.
Window
<t> Trace
Setting Basic Spectrum RF Measurement Parameters