Example:
INIT:CONT OFF
Stop the continuous sweep.
CALC:SGR:FRAM:SEL -25
Selects frame number -25.
Manual operation: See "Select Frame" on page 329
CALCulate<n>:SGRam:HDEPth <History>
CALCulate<n>:SPECtrogram:HDEPth <History>
This command defines the number of frames to be stored in the R&S FPL1000 mem-
ory.
Suffix:
<n>
.
Window
Parameters:
<History> The maximum number of frames depends on the number of
sweep points.
Range: 781 to 20000
Increment: 1
*RST: 3000
Example: //Set the history depth to 1500
CALC:SGR:SPEC 1500
Manual operation: See "History Depth" on page 409
CALCulate<n>:SGRam:LAYout <State>
CALCulate<n>:SPECtrogram:LAYout <State>
This command selects the state and size of spectrograms.
The command is available for result displays that support spectrograms.
Suffix:
<n>
.
Window
Parameters:
<State> ON
Spectrogram and trace diagram share a window.
OFF
Only the trace diagram is displayed, the spectrogram is not.
*RST: OFF
Example:
CALC4:SPEC:LAY FULL
Shows the spectrogram in window 4. The corresponding trace
diagram is hidden.
Manual operation: See "State" on page 408