User Manual

Remote Commands
R&S
®
FPL1000
830User Manual 1178.3370.02 ─ 09
CALCulate<n>:MATH<t>:MODE <Mode>
This command selects the way the R&S FPL1000 calculates trace mathematics.
Suffix:
<n>
.
Window
<t> irrelevant
Parameters:
<Mode> For more information on the way each mode works see Trace
Math Mode.
LINear
Linear calculation.
LOGarithmic
Logarithmic calculation.
POWer
Linear power calculation.
*RST: LOGarithmic
Example:
CALC:MATH:MODE LIN
Selects linear calculation.
Manual operation: See "Trace Math Mode" on page 417
CALCulate<n>:MATH<t>:POSition <Position>
This command defines the position of the trace resulting from the mathematical opera-
tion.
Suffix:
<n>
.
Window
<t> irrelevant
Parameters:
<Position> Vertical position of the trace in % of the height of the diagram
area.
100 PCT corresponds to the upper diagram border.
Range: -100 to 200
*RST: 50
Default unit: PCT
Example:
CALC:MATH:POS 100
Moves the trace to the top of the diagram area.
Manual operation: See "Trace Math Position" on page 417
CALCulate<n>:MATH<t>:STATe <State>
This command turns the trace mathematics on and off.
Suffix:
<n>
.
Window
Setting Basic Spectrum RF Measurement Parameters