User Manual

Remote Commands
R&S
®
FPL1000
840User Manual 1178.3370.02 ─ 09
n dB Down Marker................................................................................................ 881
Signal Count Marker............................................................................................. 884
Marker Demodulation............................................................................................886
Programming Examples for Using Markers and Marker Functions.......................889
11.8.8.1 Setting Up Individual Markers
The following commands define the position of markers in the diagram.
CALCulate<n>:DELTamarker<m>:AOFF...........................................................................840
CALCulate<n>:DELTamarker<m>:LINK............................................................................840
CALCulate<n>:DELTamarker<ms>:LINK:TO:MARKer<md>................................................841
CALCulate<n>:DELTamarker<m>:MODE......................................................................... 841
CALCulate<n>:DELTamarker<m>:MREFerence................................................................ 842
CALCulate<n>:DELTamarker<m>[:STATe]........................................................................ 842
CALCulate<n>:DELTamarker<m>:TRACe.........................................................................842
CALCulate<n>:DELTamarker<m>:X................................................................................. 843
CALCulate<n>:MARKer<m>:AOFF..................................................................................843
CALCulate<n>:MARKer<ms>:LINK:TO:MARKer<md>.......................................................844
CALCulate<n>:MARKer<m>[:STATe]............................................................................... 844
CALCulate<n>:MARKer<m>:TRACe................................................................................844
CALCulate<n>:MARKer<m>:X........................................................................................ 845
CALCulate<n>:DELTamarker<m>:AOFF
This command turns off all delta markers.
Suffix:
<n>
.
Window
<m> irrelevant
Example:
CALC:DELT:AOFF
Turns off all delta markers.
CALCulate<n>:DELTamarker<m>:LINK <State>
This command links delta marker <m> to marker 1.
If you change the horizontal position (x-value) of marker 1, delta marker <m> changes
its horizontal position to the same value.
Tip: to link any marker to a different marker than marker 1, use the CALCulate<n>:
DELTamarker<ms>:LINK:TO:MARKer<md> or CALCulate<n>:MARKer<ms>:
LINK:TO:MARKer<md> commands.
Suffix:
<n>
.
Window
<m> Marker
Parameters:
<State> ON | OFF | 0 | 1
Setting Basic Spectrum RF Measurement Parameters