User Manual
Remote Commands
R&S
®
FPL1000
843User Manual 1178.3370.02 ─ 09
If necessary, the command activates the marker first.
Suffix:
<n>
.
Window
<m> Marker
Parameters:
<Trace> Trace number the marker is assigned to.
Example:
CALC:DELT2:TRAC 2
Positions delta marker 2 on trace 2.
CALCulate<n>:DELTamarker<m>:X
<Position>
This command moves a delta marker to a particular coordinate on the x-axis.
If necessary, the command activates the delta marker and positions a reference
marker to the peak power.
Suffix:
<n>
.
Window
<m> Marker
Parameters:
<Position> Numeric value that defines the marker position on the x-axis.
The position is relative to the reference marker.
To select an absolute position you have to change the delta
marker mode with CALCulate<n>:DELTamarker<m>:MODE
on page 841.
A query returns the absolute position of the delta marker.
Range: The value range and unit depend on the measure-
ment and scale of the x-axis.
Default unit: HZ
Example:
CALC:DELT:X?
Outputs the absolute x-value of delta marker 1.
Manual operation: See "Marker 1/Marker 2/Marker 3/Marker 4" on page 244
See "Marker 1/Marker 2/Marker 3" on page 248
See "Marker Position X-value" on page 262
CALCulate<n>:MARKer<m>:AOFF
This command turns off all markers.
Suffix:
<n>
.
Window
<m> Marker
Example:
CALC:MARK:AOFF
Switches off all markers.
Manual operation: See "All Markers Off" on page 352
Setting Basic Spectrum RF Measurement Parameters