User Manual
Remote Commands
R&S
®
FPL1000
847User Manual 1178.3370.02 ─ 09
<m> irrelevant
Parameters:
<StepSize> STANdard
the marker moves from one pixel to the next
POINts
the marker moves from one sweep point to the next
*RST: POINts
Example:
CALC:MARK:X:SSIZ STAN
Sets the marker step size to one pixel.
Manual operation: See "Marker Stepsize" on page 354
11.8.8.3 Configuring and Performing a Marker Search
The following commands control the marker search.
CALCulate<n>:MARKer<m>:LOEXclude.......................................................................... 847
CALCulate<n>:MARKer<m>:PEXCursion.........................................................................847
CALCulate<n>:MARKer<m>:X:SLIMits[:STATe].................................................................848
CALCulate<n>:MARKer<m>:X:SLIMits:LEFT....................................................................848
CALCulate<n>:MARKer<m>:X:SLIMits:RIGHt...................................................................849
CALCulate<n>:MARKer<m>:X:SLIMits:ZOOM[:STATe]......................................................850
CALCulate<n>:THReshold..............................................................................................850
CALCulate<n>:THReshold:STATe....................................................................................850
CALCulate<n>:MARKer<m>:LOEXclude <State>
This command turns the suppression of the local oscillator during automatic marker
positioning on and off (for all markers in all windows).
Suffix:
<n>
.
irrelevant
<m> irrelevant
Parameters:
<State> ON | OFF | 0 | 1
*RST: 1
Example:
CALC:MARK:LOEX ON
Manual operation: See "Exclude LO" on page 356
CALCulate<n>:MARKer<m>:PEXCursion <Excursion>
This command defines the peak excursion (for all markers in all windows).
The peak excursion sets the requirements for a peak to be detected during a peak
search.
The unit depends on the measurement.
Setting Basic Spectrum RF Measurement Parameters