User Manual
Applications, Measurement Channels, and Result Displays
R&S
®
FPL1000
85User Manual 1178.3370.02 ─ 09
Example: Sequencer procedure
Assume the following active channel setup definition:
Tab name Application Sweep mode Sweep count
Spectrum Spectrum Cont. Sweep 5
Spectrum 2 Spectrum Single Sweep 6
Spectrum 3 Spectrum Cont. Sweep 2
IQ Analyzer IQ Analyzer Single Sweep 7
For Single Sequence, the following sweeps will be performed:
5x Spectrum, 6x Spectrum 2, 2 x Spectrum 3, 7x IQ Analyzer
For Continuous Sequence, the following sweeps will be performed:
5x Spectrum, 6x Spectrum 2, 2 x Spectrum 3, 7x IQ Analyzer,
5x Spectrum, 6x Spectrum 2, 2 x Spectrum 3, 7x IQ Analyzer,
...
For Channel-defined Sequence, the following sweeps will be performed:
5x Spectrum, 6x Spectrum 2, 2 x Spectrum 3, 7x IQ Analyzer,
5x Spectrum, 2 x Spectrum 3,
5x Spectrum, 2 x Spectrum 3,
...
Run Single/Run Cont and Single Sweep/Sweep Continuous keys
While the Sequencer is active, the [Run Single] and [Run Cont] keys control the
Sequencer, not individual sweeps. [Run Single] starts the Sequencer in single mode,
while [Run Cont] starts the Sequencer in continuous mode.
Running a Sequence of Measurements