User Manual

Remote Commands
R&S
®
FPL1000
861User Manual 1178.3370.02 ─ 09
Suffix:
<n>
.
Window
<m> Marker
CALCulate<n>:MARKer<m>:SGRam:XY:MINimum[:PEAK]
CALCulate<n>:MARKer<m>:SPECtrogram:XY:MINimum[:PEAK]
This command moves a marker to the minimum level of the spectrogram.
Suffix:
<n>
.
Window
<m> Marker
CALCulate<n>:MARKer<m>:SGRam:Y:MAXimum:ABOVe
CALCulate<n>:MARKer<m>:SPECtrogram:Y:MAXimum:ABOVe
This command moves a marker vertically to the next lower peak level for the current
frequency.
The search includes only frames above the current marker position. It does not change
the horizontal position of the marker.
Suffix:
<n>
.
Window
<m> Marker
Manual operation: See "Search Mode for Next Peak in Y-Direction" on page 358
CALCulate<n>:MARKer<m>:SGRam:Y:MAXimum:BELow
CALCulate<n>:MARKer<m>:SPECtrogram:Y:MAXimum:BELow
This command moves a marker vertically to the next lower peak level for the current
frequency.
The search includes only frames below the current marker position. It does not change
the horizontal position of the marker.
Suffix:
<n>
.
Window
<m> Marker
Manual operation: See "Search Mode for Next Peak in Y-Direction" on page 358
CALCulate<n>:MARKer<m>:SGRam:Y:MAXimum:NEXT
CALCulate<n>:MARKer<m>:SPECtrogram:Y:MAXimum:NEXT
This command moves a marker vertically to the next lower peak level for the current
frequency.
The search includes all frames. It does not change the horizontal position of the
marker.