User Manual
Remote Commands
R&S
®
FPL1000
880User Manual 1178.3370.02 ─ 09
Using Delta Markers
CALCulate<n>:DELTamarker<m>:FUNCtion:BPOWer:MODE.............................................880
CALCulate<n>:DELTamarker<m>:FUNCtion:BPOWer:RESult?...........................................880
CALCulate<n>:DELTamarker<m>:FUNCtion:BPOWer:SPAN..............................................880
CALCulate<n>:DELTamarker<m>:FUNCtion:BPOWer[:STATe]........................................... 881
CALCulate<n>:DELTamarker<m>:FUNCtion:BPOWer:MODE <Mode>
This command selects the way the results for a band power delta marker are dis-
played.
Suffix:
<n>
.
Window
<m> Marker
Parameters:
<Mode> POWer
Result is displayed as an absolute power. The power unit
depends on the CALCulate<n>:UNIT:POWer setting.
DENSity
Result is displayed as a density in dBm/Hz.
RPOWer
This setting is only available for a delta band power marker.
The result is the difference between the absolute power in the
band around the delta marker and the absolute power for the
reference marker. The powers are subtracted logarithmically, so
the result is a dB value.
[Relative band power (Delta2) in dB] = [absolute band power
(Delta2) in dBm] - [absolute (band) power of reference marker in
dBm]
For details see "Relative band power markers" on page 374.
*RST:
POWer
Manual operation: See "Power Mode" on page 376
CALCulate<n>:DELTamarker<m>:FUNCtion:BPOWer:RESult?
This command queries the results of the band power measurement.
Suffix:
<n>
.
Window
<m> Marker
Return values:
<Power> Signal power over the delta marker bandwidth.
Usage: Query only
CALCulate<n>:DELTamarker<m>:FUNCtion:BPOWer:SPAN <Span>
This command defines the bandwidth around the delta marker position.
Setting Basic Spectrum RF Measurement Parameters