User Manual
Remote Commands
R&S
®
FPL1000
898User Manual 1178.3370.02 ─ 09
11.8.9 Configuring Display Lines
The commands required to configure display lines in a remote environment are descri-
bed here.
CALCulate<n>:DLINe<dl>...............................................................................................898
CALCulate<n>:DLINe<dl>:STATe.................................................................................... 898
CALCulate<n>:FLINe<dl>...............................................................................................899
CALCulate<n>:FLINe<dl>:STATe.....................................................................................899
CALCulate<n>:PLINe<dl>...............................................................................................899
CALCulate<n>:PLINe<dl>:STATe.....................................................................................900
CALCulate<n>:TLINe<dl>...............................................................................................900
CALCulate<n>:TLINe<dl>:STATe.....................................................................................901
CALCulate<n>:DLINe<dl> <Position>
This command defines the (horizontal) position of a display line.
Suffix:
<n>
.
Window
<dl> 1 | 2
Parameters:
<Position> The value range is variable.
You can use any unit you want, the R&S FPL1000 then converts
the unit to the currently selected unit. If you omit a unit, the
R&S FPL1000 uses the currently selected unit.
*RST: (state is OFF)
Default unit: DBM
Example:
CALC:DLIN2 -20dBm
Positions the second display line at -20 dBm.
Manual operation: See "Horizontal Line 1/ Horizontal Line 2" on page 420
CALCulate<n>:DLINe<dl>:STATe <State>
This command turns a display line on and off
Suffix:
<n>
.
Window
<dl> 1 | 2
Parameters:
<State> ON | OFF | 0 | 1
OFF | 0
Switches the function off
ON | 1
Switches the function on
Example:
CALC:DLIN2:STAT ON
Turns on display line 2.