User Manual

Remote Commands
R&S
®
FPL1000
907User Manual 1178.3370.02 ─ 09
Before you can use the command, you have to select a limit line with CALCulate<n>:
LIMit<li>:NAME on page 907.
Suffix:
<n>
.
irrelevant
<li> Limit line
Parameters:
<State> ON | OFF | 0 | 1
OFF | 0
Switches the function off
ON | 1
Switches the function on
Manual operation: See "Visibility" on page 427
CALCulate<n>:LIMit<li>:LOWer:THReshold
<Threshold>
This command defines a threshold for relative limit lines.
The R&S FPL1000 uses the threshold for the limit check, if the limit line violates the
threshold.
Suffix:
<n>
.
irrelevant
<li> Limit line
Parameters:
<Threshold> Numeric value.
The unit depends on CALCulate<n>:LIMit<li>:UNIT
on page 908.
*RST: -200 dBm
Default unit: DBM
Manual operation: See "Threshold" on page 429
CALCulate<n>:LIMit<li>:NAME <Name>
This command selects a limit line that already exists or defines a name for a new limit
line.
Suffix:
<n>
.
Window
<li> Limit line
Parameters:
<Name> String containing the limit line name.
*RST: REM1 to REM8 for lines 1 to 8
Manual operation: See "Name" on page 429
Setting Basic Spectrum RF Measurement Parameters