User Manual
Remote Commands
R&S
®
FPL1000
910User Manual 1178.3370.02 ─ 09
Parameters:
<Distance> Defines the distance that the limit line moves.
The unit depends on CALCulate<n>:LIMit<li>:UNIT
on page 908.
Manual operation: See "Shift y" on page 431
CALCulate<n>:LIMit<li>:UPPer:SPACing <InterpolType>
This command selects linear or logarithmic interpolation for the calculation of an upper
limit line from one horizontal point to the next.
Suffix:
<n>
.
Window
<li> Limit line
Parameters:
<InterpolType> LINear | LOGarithmic
*RST: LIN
Manual operation: See "Y-Axis" on page 430
CALCulate<n>:LIMit<li>:UPPer:STATe <State>
This command turns an upper limit line on and off.
Before you can use the command, you have to select a limit line with CALCulate<n>:
LIMit<li>:NAME on page 907.
Suffix:
<n>
.
irrelevant
<li> Limit line
Parameters:
<State> ON | OFF | 0 | 1
OFF | 0
Switches the function off
ON | 1
Switches the function on
Manual operation: See "Visibility" on page 427
CALCulate<n>:LIMit<li>:UPPer:THReshold <Limit>
This command defines an absolute limit for limit lines with a relative scale.
The R&S FPL1000 uses the threshold for the limit check, if the limit line violates the
threshold.
Suffix:
<n>
.
irrelevant
<li> Limit line
Setting Basic Spectrum RF Measurement Parameters