User Manual
Remote Commands
R&S
®
FPL1000
914User Manual 1178.3370.02 ─ 09
Manual operation: See "Export" on page 431
11.8.10.3 Checking the Results of a Limit Check
CALCulate<n>:LIMit<li>:CLEar[:IMMediate]......................................................................914
CALCulate<n>:LIMit<li>:FAIL?........................................................................................ 914
CALCulate<n>:LIMit<li>:CLEar[:IMMediate]
This command deletes the result of the current limit check.
The command works on all limit lines in all measurement windows at the same time.
Suffix:
<n>
.
Window
<li> irrelevant
Example:
CALC:LIM:CLE
Deletes the result of the limit check.
CALCulate<n>:LIMit<li>:FAIL?
This command queries the result of a limit check in the specified window.
Note that for SEM measurements, the limit line suffix <li> is irrelevant, as only one spe-
cific SEM limit line is checked for the currently relevant power class.
To get a valid result, you have to perform a complete measurement with synchroniza-
tion to the end of the measurement before reading out the result. This is only possible
for single sweep mode.
See also INITiate<n>:CONTinuous on page 628.
Suffix:
<n>
.
Window
<li> Limit line
Return values:
<Result> 0
PASS
1
FAIL
Example:
INIT;*WAI
Starts a new sweep and waits for its end.
CALC2:LIM3:FAIL?
Queries the result of the check for limit line 3 in window 2.
Usage: Query only
Manual operation: See "Limit Check <n>" on page 177
See "Limit Check" on page 207
Setting Basic Spectrum RF Measurement Parameters