11.9.3 Storing and Loading Instrument Settings
See also:
INSTrument[:SELect] on page 622 to select the channel setup.
MMEMory:CLEar:ALL.....................................................................................................927
MMEMory:CLEar:STATe................................................................................................. 927
MMEMory:LOAD:AUTO..................................................................................................927
MMEMory:LOAD:STATe..................................................................................................928
MMEMory:LOAD:TYPE.................................................................................................. 929
MMEMory:STORe<1|2>:STATe....................................................................................... 930
MMEMory:STORe<1|2>:STATe:NEXT..............................................................................930
MMEMory:STORe<1|2>:TYPE........................................................................................ 931
SYSTem:PRESet........................................................................................................... 931
SYSTem:PRESet:CHANnel[:EXEC]................................................................................. 931
MMEMory:CLEar:ALL
This command deletes all instrument configuration files in the current directory.
You can select the directory with MMEMory:CDIRectory on page 920.
Example:
MMEM:CLE:ALL
MMEMory:CLEar:STATe <1>, <FileName>
This command deletes an instrument configuration file.
Parameters:
<1>
<FileName> String containing the path and name of the file to delete.
The string may or may not contain the file's extension.
Example:
MMEM:CLE:STAT 1,'TEST'
MMEMory:LOAD:AUTO <1>, <FileName>
This command restores an instrument configuration and defines that configuration as
the default state.
The default state is restored after a preset (*RST) or after you turn on the
R&S FPL1000.
Parameters:
<1>
<FileName> 'Factory'
Restores the factory settings as the default state.
'<file_name>
String containing the path and name of the configuration file.
Note that only instrument settings files can be selected for the
startup recall function; channel setup files cause an error.