11.10 Configuring the R&S FPL1000
The remote commands required to set up the R&S FPL1000 are described here.
● Configuring the Reference Frequency.................................................................. 955
● Calibration and Checks.........................................................................................955
● Working with Transducers.....................................................................................958
● Customizing the Screen Layout............................................................................ 962
● Remote Commands for Language Settings..........................................................967
● Configuring the Network and Remote Control...................................................... 967
● Checking the System Configuration......................................................................971
● Using Service Functions....................................................................................... 975
11.10.1 Configuring the Reference Frequency
[SENSe:]ROSCillator:SOURce........................................................................................ 955
[SENSe:]ROSCillator:SOURce <Source>
This command selects the reference oscillator.
If you want to select the external reference, it must be connected to the R&S FPL1000.
Parameters:
<Source> INTernal
The internal reference is used (10 MHz)
EXTernal
The external reference from the "REF INPUT 10 MHZ" connec-
tor is used; if none is available, an error flag is displayed in the
status bar
Example:
ROSC:SOUR EXT
11.10.2 Calibration and Checks
The following commands control calibration and checks on the R&S FPL1000.
CALibration[:ALL]...........................................................................................................956
CALibration:RESult?.......................................................................................................956
DIAGnostic:SERVice:INPut:PULSed:CFRequency............................................................ 956
DIAGnostic:SERVice:INPut:PULSed:MCFRequency..........................................................957
DIAGnostic:SERVice:INPut:RF[:SPECtrum]......................................................................957
Configuring the R&S