User Manual

Remote Commands
R&S
®
FPL1000
957User Manual 1178.3370.02 ─ 09
Before you can use the command, you have to feed in a calibration signal with
DIAGnostic:SERVice:INPut[:SELect] on page 957.
Parameters:
<Frequency> Possible frequencies of the calibration signal are fix.
If you define a frequency that is not available, the R&S FPL1000
uses the next available frequency. Example: a frequency of
20 MHz is rounded up to the next available frequency (25 MHz).
*RST: 50 MHz
Default unit: Hz
Manual operation: See "Calibration Frequency RF" on page 520
DIAGnostic:SERVice:INPut:PULSed:MCFRequency
<Frequency>
This command sets the calibration frequency for frequencies greater than 7 GHz. This
command only takes effect if a microwave calibration signal is selected for input
(DIAGnostic:SERVice:INPut[:SELect] on page 957).
Parameters:
<Frequency> *RST: 7 GHz
Default unit: Hz
Example:
DIAG:SERV:INP:PULS:MCFR 7,1 GHz
DIAGnostic:SERVice:INPut:RF[:SPECtrum] <Bandwidth>
This command selects the bandwidth of the calibration signal.
Parameters:
<Bandwidth> NARRowband | BROadband
NARRowband
Narrowband signal for power calibration of the frontend.
BROadband
Broadband signal for calibration of the IF filter.
DIAGnostic:SERVice:INPut[:SELect] <Signal>
This command activates or deactivates the use of an internal calibration signal as input
for the R&S FPL1000.
Parameters:
<Signal> CALibration
Uses the calibration signal as RF input.
RF
Uses the signal from the RF input.
*RST: RF
Example:
DIAG:SERV:INP CAL
Uses the calibration signal as RF input.
Configuring the R&S FPL1000