User Manual
Remote Commands
R&S
®
FPL1000
962User Manual 1178.3370.02 ─ 09
String Unit
'DBUV/M' dBμV/m
(Requires R&S FPL1-K54 (EMI measurements) option.)
'DBUA' dBμA
'DBUA/M' dBμA/m
(Requires R&S FPL1-K54 (EMI measurements) option.)
'DBPW' dBpW
'DBPT' dBpT
MMEMory:LOAD<n>:TFACtor <FileName>
Loads the transducer factor from the selected file in .CSV format.
Suffix:
<n>
.
irrelevant
Parameters:
<FileName> String containing the path and name of the CSV import file.
Example:
MMEM:LOAD:TFAC 'C:\TEST.CSV'
Manual operation: See "Import" on page 530
MMEMory:STORe<n>:TFACtor <FileName>, <TransdName>
This command exports transducer factor data to an ASCII (CSV) file.
For details on the file format see Chapter 9.7.3, "Reference: Transducer Factor File
Format", on page 531.
Suffix:
<n>
.
irrelevant
Parameters:
<FileName> Name of the transducer factor to be exported.
<TransdName> Name of the transducer factor to be exported.
Example:
MMEM:STOR:TFAC 'C:\TEST', 'Transducer1'
Stores the transducer factor named "Transducer1" in the file
TEST.CSV.
Manual operation: See "Export" on page 530
11.10.4 Customizing the Screen Layout
The remote commands required to set up the display of the R&S FPL1000 are descri-
bed here.
● General Display Settings and Items......................................................................963
● Colors and Themes...............................................................................................966
Configuring the R&S
FPL1000