Remote commands exclusive to configuring a network and remote control
SYSTem:COMMunicate:GPIB[:SELF]:ADDRess............................................................... 968
SYSTem:COMMunicate:GPIB[:SELF]:RTERminator.......................................................... 968
SYSTem:DISPlay:LOCK................................................................................................. 969
SYSTem:DISPlay:UPDate...............................................................................................969
SYSTem:ERRor:DISPlay................................................................................................ 969
SYSTem:IDENtify:FACTory..............................................................................................969
SYSTem:IDENtify[:STRing]............................................................................................. 969
SYSTem:KLOCk............................................................................................................ 970
SYSTem:LANGuage.......................................................................................................970
SYSTem:LXI:INFO......................................................................................................... 970
SYSTem:LXI:LANReset.................................................................................................. 971
SYSTem:LXI:MDEScription............................................................................................. 971
SYSTem:LXI:PASSword..................................................................................................971
SYSTem:REVision:FACTory............................................................................................ 971
SYSTem:COMMunicate:GPIB[:SELF]:ADDRess <Address>
This command sets the GPIB address of the R&S FPL1000.
Parameters:
<Address> Range: 0 to 30
*RST: (no influence on this parameter, factory default 20)
Example:
SYST:COMM:GPIB:ADDR 18
Manual operation: See "GPIB Address" on page 582
SYSTem:COMMunicate:GPIB[:SELF]:RTERminator <Terminator>
This command selects the GPIB receive terminator.
Output of binary data from the instrument to the control computer does not require
such a terminator change.
Parameters:
<Terminator> LFEOI | EOI
LFEOI
According to the standard, the terminator in ASCII is <LF>
and/or <EOI>.
EOI
For binary data transfers (e.g. trace data) from the control com-
puter to the instrument, the binary code used for <LF> might be
included in the binary data block, and therefore should not be
interpreted as a terminator in this particular case. This can be
avoided by using only the receive terminator EOI.
*RST: LFEOI
Example:
SYST:COMM:GPIB:RTER EOI
Manual operation: See "GPIB Terminator" on page 583