User Manual
Remote Commands
R&S
®
FPL1000
969User Manual 1178.3370.02 ─ 09
SYSTem:DISPlay:LOCK <State>
Defines whether the "Display Update" function remains available in remote operation
or not.
Parameters:
<State> ON | OFF | 0 | 1
OFF | 0
The function remains available.
ON | 1
The function is not available and the display is not updated dur-
ing remote operation.
*RST: 0
SYSTem:DISPlay:UPDate <State>
This command turns the display during remote operation on and off.
If on, the R&S FPL1000 updates the diagrams, traces and display fields only.
The best performance is obtained if the display is off during remote control operation.
Parameters:
<State> ON | OFF | 1 | 0
*RST: 0
Example:
SYST:DISP:UPD ON
Manual operation: See "Remote Display Update" on page 583
SYSTem:ERRor:DISPlay <State>
This command the error display during remote operation on and off.
If activated, the R&S FPL1000 displays a message box at the bottom of the screen that
contains the most recent type of error and the command that caused the error.
Parameters:
<State> ON | OFF | 1 | 0
*RST: 0
Example:
SYST:ERR:DISP ON
Manual operation: See "Display Remote Errors" on page 583
SYSTem:IDENtify:FACTory
This command resets the query to *IDN? to its default value.
Manual operation:
See "Reset to Factory String" on page 582
SYSTem:IDENtify[:STRing] <String>
This command defines the response to *IDN?.
Configuring the R&S