User Manual
Remote Commands
R&S
®
FPL1000
970User Manual 1178.3370.02 ─ 09
Parameters:
<String> String containing the description of the instrument.
Manual operation: See "Identification String" on page 582
SYSTem:KLOCk <State>
This command activates the local lockout (remote control) or returns to the local mode.
Parameters:
<State> ON
LLO (local lockout)
OFF
GTL (go to local)
*RST: OFF
Example:
SYST:KLOC ON
Activates LLO (remote control)
Manual operation: See "Local" on page 590
SYSTem:LANGuage <Language>
This function is used to emulate previous R&S signal and spectrum analyzers.
Parameters:
<Language> "FSV" | "SCPI"
*RST: SCPI
Example:
SYST:LANG FSV
The commands and settings for the R&S FSV instrument are
used.
SYSTem:LXI:INFO
This command queries the LXI settings.
Return values:
<LXIInfo> <current version> | <LXI class> | <Computername> |
<MAC adress> | <IP adress> | <Auto MDIX>
String containing the current LXI parameters.
• <version>
• <LXIClass>
• <ComputerName>
• <MACAddress>
• <IPAddress>
• <AutoMDIX>
Manual operation: See "Current LXI Configuration" on page 587
Configuring the R&S FPL1000