User Manual
Common Instrument Functions
R&S
®
FPL1000
159User Manual 1178.3370.02 ─ 03
ter indicates a possible deviation. The current temperature of the RF frontend can be
queried using a remote command (see SOURce:TEMPerature:FRONtend?
on page 819).
Touchscreen alignment
When the device is delivered, the touchscreen is initially calibrated. However, to
ensure that the touchscreen responds to the finger contact correctly, a touchscreen
alignment is required.
Alignment of the touchscreen is useful:
●
At first use
●
After an image update or after exchanging a hard disk
●
If you notice that touching a specific point on the screen does not achieve the cor-
rect response
●
If the position of the instrument has been changed and you cannot look straight on
the screen
●
If another person operates the instrument
6.3.7.2 Alignment Settings
Access: SETUP > "Alignment"
Both the instrument and the touchscreen can be aligned when necessary (see Chap-
ter 6.3.7.1, "Basics on Alignment", on page 158).
Start Self Alignment ................................................................................................... 160
Abort Self Alignment .................................................................................................. 160
Await Warm-Up Operation before Self Alignment.......................................................160
Shut down Device after Self Alignment.......................................................................160
Starting Touch Screen Alignment .............................................................................. 160
Alignment Results: ..................................................................................................... 160
General Instrument Setup