User Manual

Common Instrument Functions
R&S
®
FPL1000
203User Manual 1178.3370.02 ─ 03
Remote command:
SYSTem:IDENtify:FACTory on page 832
Remote Display Update
Defines whether the display of the R&S FPL1000 is updated when changing from man-
ual operation to remote control.
Turning off the display update function improves performance during remote control.
Remote command:
SYSTem:DISPlay:UPDate on page 832
GPIB Terminator
Changes the GPIB receive terminator.
"LFEOI"
According to the standard, the terminator in ASCII is <LF> and/or
<EOI>.
"EOI"
For binary data transfers (e.g. trace data) from the control computer
to the instrument, the binary code used for <LF> might be included in
the binary data block, and therefore should not be interpreted as a
terminator in this particular case. This can be avoided by using only
the receive terminator EOI.
Remote command:
SYSTem:COMMunicate:GPIB[:SELF]:RTERminator on page 831
I/O Logging
Activates or deactivates the SCPI error log function. All remote control commands
received by the R&S FPL1000 are recorded in the following log file:
C:\ProgramData\Rohde-Schwarz\ZNL-FPL\ScpiLogging\ScpiLog.txt
Logging the commands may be extremely useful for debug purposes, e.g. in order to
find misspelled keywords in control programs.
Remote command:
SYSTem:CLOGging on page 525
Display Remote Errors
Activates and deactivates the display of errors that occur during remote operation of
the R&S FPL1000. If activated, the R&S FPL1000 displays a message box at the bot-
tom of the screen that contains the type of error and the command that caused the
error.
The error message remains in place when you switch to "Local" mode. To close the
message box, select the "Close" icon.
Only the most recent error is displayed in remote mode. However, in local mode, all
errors that occurred during remote operation are listed in a separate tab of the "Net-
work + Remote" dialog box (see Chapter 6.4.2.5, "Remote Errors", on page 207).
Network and Remote Operation