CALC:MARK:FUNC:POW:RES? OBW, see CALCulate<n>:MARKer<m>:FUNCtion:
POWer<sb>:RESult? on page 538
7.2.5.3 OBW Configuration
Access: "Overview" > "Select Measurement" > "OBW" > "OBW Config"
This measurement is not available in zero span.
Configuring search limits for OBW measurement
The OBW measurement uses the same search limits as defined for marker search
(see " Search Limits " on page 432). However, only the left and right limits are consid-
ered.
The remote commands required to perform these tasks are described in Chapter 8.6.5,
"Measuring the Occupied Bandwidth", on page 560.
% Power Bandwidth ...................................................................................................284
Channel Bandwidth ....................................................................................................285
Adjust Settings ........................................................................................................... 285
Search Limits ( Left / Right )........................................................................................285
Deactivating All Search Limits ....................................................................................285
% Power Bandwidth
Defines the percentage of total power in the displayed frequency range which defines
the occupied bandwidth. Values from 10 % to 99.9 % are allowed.
Remote command:
[SENSe:]POWer:BANDwidth on page 560