User Manual
The Spectrum Application (RF Measurements)
R&S
®
FPL1000
304User Manual 1178.3370.02 ─ 03
VBW
Sets the video bandwidth for this range.
For details on the VBW, see Chapter 7.6.1.2, "Smoothing the Trace Using the Video
Bandwidth", on page 391.
Remote command:
[SENSe:]ESPectrum<sb>:RANGe<range>:BANDwidth:VIDeo on page 567
Sweep Time Mode
Activates or deactivates the auto mode for the sweep time.
Currently, only auto mode is available for the R&S FPL1000.
For details on the sweep time mode, see Chapter 7.6.1.7, "How Long the Data is Mea-
sured: Sweep Time ", on page 394
Remote command:
[SENSe:]ESPectrum<sb>:RANGe<range>:SWEep:TIME:AUTO on page 578
Sweep Time
Sets the sweep time value for the range.
For details on the sweep time, see Chapter 7.6.1.7, "How Long the Data is Measured:
Sweep Time ", on page 394
Remote command:
[SENSe:]ESPectrum<sb>:RANGe<range>:SWEep:TIME on page 577
Ref Level
Sets the reference level for the range.
For details on the reference level, see Chapter 7.5.1.1, "Reference Level",
on page 382.
Remote command:
[SENSe:]ESPectrum<sb>:RANGe<range>:RLEVel on page 577
RF Att Mode
Activates or deactivates the auto mode for RF attenuation.
For details on attenuation, see Chapter 7.5.1.2, "RF Attenuation", on page 383.
Remote command:
[SENSe:]ESPectrum<sb>:RANGe<range>:INPut:ATTenuation:AUTO
on page 569
RF Attenuation
Sets the attenuation value for the range.
Remote command:
[SENSe:]ESPectrum<sb>:RANGe<range>:INPut:ATTenuation on page 569
Preamp
Switches the preamplifier on or off.
For details on the preamplifier, see " Preamplifier " on page 387.