User Manual

The Spectrum Application (RF Measurements)
R&S
®
FPL1000
319User Manual 1178.3370.02 ─ 03
5. In the file selection dialog box, select a storage location and filename for the result
file.
6. Select the "Save" button.
The file with the specified name and the extension .dat is stored in the defined
storage location.
7.2.6.7 Measurement Example: Multi-SEM Measurement
The following measurement example demonstrates an SEM measurement for a signal
with multiple sub blocks.
A programming example demonstrating a SEM measurement in a remote environment
is provided in Chapter 8.6.6.10, "Example: SEM Measurement", on page 587.
Test setup:
Signal generator settings (e.g. R&S FPL1000 SMW):
Device Standard Center frequency Level Test model
SigGen 1 3GPP/FDD 900 MHz 0 dBm 1-16
SigGen 2 EUTRA/LTE 906.5 MHz 0 dBm 1_1_5MHz
Setting up the measurement
1. Preset the R&S FPL1000.
2. Set the center frequency to 903.25 MHz.
3. Set the reference level to 10 dBm with an offset of to 30 dB.
4. Press the MEAS key or select "Select Measurement" in the "Overview" .
5. Select the "SEM" measurement function.
Measurements and Results