User Manual

The Spectrum Application (RF Measurements)
R&S
®
FPL1000
325User Manual 1178.3370.02 ─ 03
Child node Attribute Value Parameter description Mand.
WeightingFilter
Only if Referen-
cePower method
is TX Channel
Power and the
range is the ref-
erence range
Type
RRC | CFilter Type of the weighting filter Yes
Roll Off Factor
0…1 Excess bandwidth of the fil-
ter
Only if the filter
type is RRC
Bandwidth
<bandwidth in Hz> Filter bandwidth Only if the filter
type is RRC
FrequencyRange
Yes
Start
<frequency in Hz> Start value of the range Yes
Stop
<frequency in Hz> Stop value of the range Yes
Limit
dBm/Hz | dBm |
dBc | dBr | dB
A Range must contain
exactly two limit nodes; one
of the limit nodes has to
have a relative unit (e.g.
dBc), the other one must
have an absolute unit (e.g.
dBm)
Yes
Start Value <numeric_value> Power limit at start fre-
quency
Yes
Unit dBm/Hz | dBm |
dBc | dBr | dB
Sets the unit of the start
value
Stop Value <numeric_value> Power limit at stop fre-
quency
Unit dBm/Hz | dBm |
dBc | dBr | dB
Sets the unit of the stop
value
LimitFailMode
Absolute | Relative
| Absolute and Rel-
ative | Absolute or
Relative
If used, it has to be identical
to DefaultLimitFailMode
No
RBW Bandwidth <bandwidth in Hz> " RBW " on page 303 Yes
Type NORM | PULS |
CFIL | RRC
No
VBW Bandwidth <bandwidth in Hz> " VBW " on page 304 Yes
Detector
NEG | POS |
SAMP | RMS |
AVER | QUAS
If used, it has to be identical
in all ranges.
No
Sweep Mode Manual | Auto " Sweep Time Mode "
on page 304
Yes
Time <time in sec> " Sweep Time "
on page 304
No
Amplitude
No
Measurements and Results