tions of the absolute power from the limit line for each range, and the results can be
checked against defined limits automatically.
7.2.7.2 Spurious Emissions Measurement Results
The measured signal, including any spurious emissions, and optionally the detected
peaks are displayed in the Spurious Emissions measurement diagram. If defined, the
limit lines and the limit check results are also indicated. In addition to the graphical
results, a result table can be displayed to evaluate the measured powers and limit
check results (see also "Limit Lines in Spurious Measurements" on page 330). The
details of the evaluation list can be configured.
The following information is provided in the evaluation list for each range:
Column Description
Range Low Frequency range start for the range the peak value belongs to
Range Up Frequency range end for the range the peak value belongs to
RBW RBW of the range
Frequency Frequency at the peak value
Power Abs Absolute power level at the peak value
ΔLimit Deviation of the absolute power level from the defined limit for the peak
value
By default, one peak per range is displayed. However, you can change the settings to:
●
Display all peaks
Measurements and Results